| Model Number |
701116090 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
A facility reported that during a bone biopsy procedure, the needle became totally disconnected from the hub during the acquisition of a sample.This resulted in a large blood spray covering the physician and equipment.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6.And h.10.After three notifications, no product was returned for evaluation.An image of the device was provided.The image showed the device split distally, and the complaint was confirmed.Without the actual device the examine, a definitive root cause could not be determined.A review of the batch records confirmed that the product was manufactured according to specification and no deviations or anomalies were found.Argon is currently exploring options to mitigate the possiblity of recurrence of this issue.If the sample is returned at a future date, this complaint may be reopened for further evaluation at that time.There have been no other complaints reported in the lot.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|