This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, it was determined, that an abnormality of the device was not the cause of the reported event.However, the root cause could not be identified.The event can be detected/prevented, by following the instructions for use (ifu), which state: ¿use this instrument in an environment equipped to accommodate open surgery.And have the hospitalization plan prepared in case any problem occurs, that may not be resolved endoscopically¿.¿this instrument assumes, that it is used only when the following conditions are met, but it should be noted, that in certain cases, use of this instrument could lead to a surgical operation, perforation or bleeding.Keep this possibility in mind when using this instrument¿.¿this instrument has been designed for use by a physician.The operator of this instrument should deepen the understanding of the clinical procedures by referring to video and other medical materials showing clinical cases.And by receiving sufficient training in training facilities¿.¿this instrument has been designed for use under situations in which, surgical operation is available as an emergency measure¿.¿if you use the instrument several times, during one procedure.Confirm, that there is no irregularity with its operation and appearance, each time you use it.Do not use the instrument if you detect any abnormality.The breakage of the distal end such as the operation wire¿s detachment could cause mucous membrane damages¿.¿be sure to check the output power of the electrosurgical unit before use.If the unit is used without prior output setting, perforation, bleeding or mucous membrane damage may result¿.¿always operate the electrosurgical unit at the minimum output level.And for the minimum time necessary to successfully complete the procedures.Excessive output level and time may result in patient injury, such as perforation, bleeding or mucous membrane damage¿.¿do not apply current, while pressing the instrument against tissue, while the forceps are closed or opened.This may cause patient injury such as perforation, massive bleeding or tissue damage¿.This supplemental report includes a correction to g2 and h6 (type of investigation) from the initial medwatch.Olympus will continue to monitor field performance for this device.
|