| Catalog Number |
510035VER |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A supplemental report will be submitted once the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that when attempting a renal angiography after successful cannulation of the renal ostium they were manipulating/torquing and positioning the catheter for a contrast injection.During the contrast hand injection, the catheter tip started to detach.The account was able to rewire the catheter with a guidewire so that the guidewire and catheter tip were successfully removed from this patient as a unit.After catheter removal ex-vivo, they continued to manipulate the catheter again checking for catheter flexibility when the tip fractured.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|