| Catalog Number |
4540018-07 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report (b)(4): the complaint is under evaluation.A follow-up report will be provided after the examination results are available.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility / translation of user facility information by bbm sales organization in switzerland: two patients were admitted to the department and therapy was started for the one on (b)(6) 2024 at 3 p.M., and for other around 4:35 p.M.Therapies ended on (b)(6) 2024 around 10 pm and the patients were quite red.The therapy ended up to 10 hours before the scheduled time.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).Root cause analysis: device history record (dhr): reviewed the dhr for batch 23f02ged71, there is no abnormality and no such defect detected at in process and at final control inspection.Sample/s evaluation: the flow rate report of affected batch 23f02ged71 was reviewed.For final control flow rate (after eto) report, the average flow rate deviation from the nominal flow rate was between 4.38% and 8.75%.Received 1 easypump ii lt 400-40-s-eu/sa, which is a reference sample as bbm received it in original packaging.As received condition, the clamp clip of received sample was clamped, and wing cap was connected to the patient connector.Visual inspection had done throughout the received sample.No abnormalities was observed.The sample was decontaminated and sent for flow rate test (202403041658).The flow rate deviation from nominal flow rate for received sample was -0.91%.The flow rate of received sample was within the specification ±15% deviation from nominal flow rate.Summary of root cause analysis: since the flow rate of received sample was within the specification, hence we considered this complaint as not confirmed.Cause : cause could not be determine.The flow rate of received sample was within the specification after sent for flow rate test, which is according to test method 116008 under lab condition.Corrections/containment plans with effective date: not applicable.Corrective actions with effective date: not applicable.Justification: not confirmed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
| |
|
Search Alerts/Recalls
|