• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is confirmed and was determined to be use related.Five 22g x 3/4" powerloc max infusion sets were returned for evaluation.The returned product samples were evaluated and a tear in the extension tube was seen at the interface of the proximal luer adapter.The fracture surfaces of the damage contained striation-like patterns which were indicative of flexural fatigue based failures, which may have been a contributing factor to the extension tubing damage.Repetitive mechanical stresses such as twisting and kinking may have contributed to the observed event; however, it appeared that additional unidentified factors also contributed.The device is a supplied component and the supplier has been notified of this event.This complaint will be recorded for future trending and monitoring purposes.H3 other text : evaluation findings are in section h11.
 
Event or Problem Description
It was reported that customer started seeing a trend with the port a cath needle (22g x 0.75 in.): repeated instances of cracked tubing with leaking.A specific number of affected devices or patients was not provided by the complainant.Five devices were returned for evaluation.This report addresses the third device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18794547
Report Number3006260740-2024-00798
Device Sequence Number12602782
Product Code FPA
UDI-Device Identifier00801741047497
UDI-Public(01)00801741047497
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2025
Device Model NumberN/A
Device Catalogue Number0142275
Device Lot NumberASGXFC025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2024
Initial Report FDA Received Date02/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-