The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the insertion flexible tube buckled, the electrical pin connector corroded, the nozzle gluing missing, the u/d and the r/l knobs disinfections damage, the lcb (light carrying bundle) defective, the operation channel (primary) resistance, the bending rubber discolored, and the operation channel (primary) kink; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|