• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - PARTIALLY COVERED; ESW PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - PARTIALLY COVERED; ESW PROSTHESIS, ESOPHAGEAL Back to Search Results
Catalog Number EVO-20-25-10-E
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Device issue: inadequate stent expansion.Relationship questions study device: blank, procedure: not related.Pre-existing: blank.Device deficiency: blank as per attached patient casebook, " 1.Date of procedure: (b)(6) 2018.2.Reason for stent placement: malignant obstruction (extrinsic).3.Morphology of study lesion: length of study lesion: 8 (cm); diameter of esophagus at the location of the study lesion 9 (mm) "."post-procedural information: 4.Was tumor-reduction therapy documented during follow up? no.5.Were there any medical problems or complications documented after the study procedure and through follow-up medical record review? no.6.Were any device deficiencies documented after the study procedure and through follow-up medical record review; including device deficiencies that may have occurred during stent repositioning or removal? no"."2.Adverse event onset (days post stent placement): 26 7.Were any adverse events documented during the stent procedure? yes.8.Were any device deficiencies documented during the placement procedure? no.3.Event category (select only one event per form): device issues -- inadequate stent expansion.4.Event treatment: endoscopic - study stent repositioned.5.Was the event considered to be related to the study procedure? not related".This file will capture user error of the stent being repositioned 26 days post placement.Patient outcome; treatment: endoscopic; study stent repositioned.
 
Additional Manufacturer Narrative
Pma/510(k) # k162717.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Cancellation report is being submitted due to the receipt of clarification from the product manager that there was no case of user error.It is likely that the stent migrated due to the inadequate stent expansion reported, therefore the customer would have had no choice but to reposition the stent.This complaint is no longer required.Reference pr # (b)(4)/ mdr (b)(4) device issue: inadequate stent expansion.
 
Additional Manufacturer Narrative
Pma/510(k) # k162717 cancellation report is being submitted due to the receipt of clarification from the product manager that there was no case of user error.It is likely that the stent migrated due to the inadequate stent expansion reported, therefore the customer would have had no choice but to reposition the stent.This complaint is no longer required.Reference pr # (b)(4)/ mdr (b)(4) device issue: inadequate stent expansion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - PARTIALLY COVERED
Common Device Name
ESW PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key18795537
Report Number3001845648-2024-00076
Device Sequence Number7130948
Product Code ESW
UDI-Device Identifier10827002480312
UDI-Public(01)10827002480312(17)200320(10)C1465258
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2020
Device Catalogue NumberEVO-20-25-10-E
Device Lot NumberC1465258
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer02/01/2024
Initial Report FDA Received Date02/28/2024
Supplement Report FDA Received Date07/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/20/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight41 KG
-
-