| Catalog Number |
EVO-20-25-10-E |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device issue: inadequate stent expansion.Relationship questions study device: blank, procedure: not related.Pre-existing: blank.Device deficiency: blank as per attached patient casebook, " 1.Date of procedure: (b)(6) 2018.2.Reason for stent placement: malignant obstruction (extrinsic).3.Morphology of study lesion: length of study lesion: 8 (cm); diameter of esophagus at the location of the study lesion 9 (mm) "."post-procedural information: 4.Was tumor-reduction therapy documented during follow up? no.5.Were there any medical problems or complications documented after the study procedure and through follow-up medical record review? no.6.Were any device deficiencies documented after the study procedure and through follow-up medical record review; including device deficiencies that may have occurred during stent repositioning or removal? no"."2.Adverse event onset (days post stent placement): 26 7.Were any adverse events documented during the stent procedure? yes.8.Were any device deficiencies documented during the placement procedure? no.3.Event category (select only one event per form): device issues -- inadequate stent expansion.4.Event treatment: endoscopic - study stent repositioned.5.Was the event considered to be related to the study procedure? not related".This file will capture user error of the stent being repositioned 26 days post placement.Patient outcome; treatment: endoscopic; study stent repositioned.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # k162717.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
Cancellation report is being submitted due to the receipt of clarification from the product manager that there was no case of user error.It is likely that the stent migrated due to the inadequate stent expansion reported, therefore the customer would have had no choice but to reposition the stent.This complaint is no longer required.Reference pr # (b)(4)/ mdr (b)(4) device issue: inadequate stent expansion.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # k162717 cancellation report is being submitted due to the receipt of clarification from the product manager that there was no case of user error.It is likely that the stent migrated due to the inadequate stent expansion reported, therefore the customer would have had no choice but to reposition the stent.This complaint is no longer required.Reference pr # (b)(4)/ mdr (b)(4) device issue: inadequate stent expansion.
|
| |
|