| Model Number |
560BCS1 |
| Medical Device Problem Codes |
Material Deformation (2976); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation:the reported motor failure was verified during service.Service technician carried out general cleaning and visual inspection.The system manager board, and batteries were replaced.Service technician performed calibrations for manufacturing parameters and performed power supply, system manager, voltage validation,battery and charger voltages, pressure ports, and touchscreen checks.Service technician verified parameters and functions by hyperterminal.The internal failure was evident in the 560a motor during tests, the latter needs to be replaced with a new one before service can continue calibrating rpm and flow.Post repair testing will be performed per specifications.Note that the instrument was not returned to medtronic facility but was serviced by field service technician medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a bio-console instrument, it was reported that it had a failed motor and damaged component when motor was connected.The use of the instrument was unknown.There was no adverse patient effect associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received information that prior to use of a bio-console instrument, it was reported that it had a failed motor and damaged component when motor was connected.The use of the instrument was unknown.There was no patient involvement, so no adverse effect o ccurred.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|