DEPUY MITEK LLC US RIGIDFIX FEM 3.3MM S/T XPIN; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 210133
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: the device manufacture date is currently unavailable.(b)(4) as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported by the healthcare professional in china that during an anterior cruciate ligament repair procedure on (b)(6) 2024, it was observed that the rigidfix fem 3.3mm s/t xpin device was found to be bent and breakage.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the investigation has been updated to reflect the correct information: investigation summary according to the information received, it was reported that during the surgery, the rigidfix was found to be bent and breakage(as the photo shows.No additional information could be provided.¿ ¿the product has not returned to depuy synthes mitek, however photos were provided for review.See attachments (¿¿.Jpg, ¿¿¿.Jpg) ¿ the photo investigation revealed that two rigidfix fem 3.3mm s/t xpin were shown.One device was broken at the medial part of the device, the other pin does not show any anomalies.The overall complaint was confirmed as the observed condition of the rigidfix fem 3.3mm s/t xpin would contribute to the complained device issue.¿ based on the investigation findings, the potential cause is traced to a procedural variable, such handling of the device or product interaction during procedure.As per ifu- 104144, use instructions are provided, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX FEM 3.3MM S/T XPIN
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 
3035526892
MDR Report Key18795905
Report Number1221934-2024-00594
Device Sequence Number17949632
Product Code HTY
UDI-Device Identifier10886705000890
UDI-Public10886705000890
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K974341
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number210133
Device Lot Number141L226
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2024
Supplement Date Received by Manufacturer02/29/2024
02/29/2024
Initial Report FDA Received Date02/28/2024
Supplement Report FDA Received Date03/01/2024
03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-