| Catalog Number |
9-PDA-004 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that a 6-4mm amplatzer duct occluder was selected for implant on(b)(6)2024 using an unknown abbott delivery system.During implant the left disc of the device took on a cobra shape.The device was removed from the patient and replaced with a non-abbott device.There were no interactions with cardiac structures.There were no angulations or kinks noted in the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.The patient status was stable.There were no adverse effects to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity during implant was reported.The investigation confirmed the device met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|