| Catalog Number |
9-ASD-034 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 34mm amplatzer septal occluder was chosen for implantation utilizing a non-abbott 14f ( 4.67mm inner diameter) delivery sheath.During deployment the device deformed into a cobra shape.There was no kink or bend in the delivery system and no interaction with cardiac structures.The device was removed.No replacement was implanted.Patient was reported to be stable.No patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of cobra device deformation was reported.The device was returned to abbott for investigation and the functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|