Model Number A-TCSE-F
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Endocarditis (1834); Perforation of Esophagus (2399)
Date of Event 01/15/2024
Type of Reportable Event Death
Event or Problem Description
The patient underwent an atrial fibrillation redo procedure on (b)(6) 2024.The patient presented with symptoms including vomiting, headache, and loss of consciousness.The patient died on (b)(6) 2024.The physician believes the cause of death was mitral endocarditis resulting from a possible atrio esophageal fistula.The official cause of death is reported as mitral endocarditis on native valve with cerebral, renal, splenic and coronary emboli with a positive blood culture for veillonella rogosae anaerobic organisms.It is noted that on the day of the procedure, diagnostic testing was done with ultrasound imaging and scanner and there was no evidence of endocarditis.Endocarditis was also not a pre-existing condition, the patient's medical history included rhythmic cardiomyopathy due to atrial fibrillation.There were no reported performance issues with any abbott device.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis; however, the log files from the tactisys quartz system were returned.The log file analysis concluded that the tacticath catheter performed as intended.The optical fibers met specifications, the recorded temperatures indicated cooling during rf ablation, and contact force measurements were displayed throughout the duration of the log files.Baseline was noted to not be reset after insertion and reinsertion into the patient as recommended in the tacticath se contact force ablation catheter instructions for use (ifu); however, due to unknown procedural conditions we are unable to conclusively determine the cause of the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Event or Problem Description
This report includes updated device information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18799168
Report Number3008452825-2024-00101
Device Sequence Number7342565
Product Code OAE
UDI-Device Identifier05415067027696
UDI-Public(01)05415067027696(10)9254116(17)250228
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P130026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberA-TCSE-F
Device Lot Number9254116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2024
Supplement Date Received by Manufacturer03/28/2024
Initial Report FDA Received Date02/28/2024
Supplement Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-