| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that they have identified a communication problem between mosaiq and the cyberknife control console.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the product and the reported information.According to the treatment records, 3,713.6 mu was given to the patient of the 8,852.3 mu prescribed.The 3,713.6 mu was recorded in mosaiq, leaving a remainder of 5,138.7 mu.The software logs indicate that the remainder may have been selected in mosaiq for delivery, but no treatment records were received for the completion of the partial treatment.The remainder of 5,138.7 mu was later recorded manually by the user.There is no evidence of patient mistreatment.Mosaiq did not have any malfunction and is working as designed and intended.
|
| |
|
Event or Problem Description
|
|
The customer reported that they have identified a communication problem between mosaiq and the cyberknife control console.
|
| |
|