| Catalog Number |
963206000 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212); Fracture (1260)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Insufficient Information (4580)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the setting instrument for the medullary canal stopper broke off when it was unscrewed.No force was used.The approximately 6 cm long fragment could not be removed and had to remain in the patient.
|
| |
|
Event or Problem Description
|
|
Additional information was received and stated that the operation was delayed by half an hour.They could not remove the broken part and pushed it into the medullary space distal.The patient now has an additional foreign body and the operation and anesthesia took longer.It was also reported that the instrument broke into 2 parts.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, d4 (lot), d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, hcp is reporting the setting instrument for the medullary canal stopper broke off when it was unscrewed.No force was used.The approximately 6 cm long fragment could not be removed and had to remain in the patient.The product was not returned to depuy synthes, however photos were provided for review.See attachment (b)(4) photos received from customer.The photo investigation revealed that the hardinge fem cement rester was found broken from the shaft, the fracture originated from the number 7 marked on the shaft, indicating that approximately 6 cm from the tip broke off the instrument.No postoperative x-rays were provided to confirm the foreign body left in patient.The shaft breakage is consistent with the user applying excessive leverage/torque on the hardinge cem rester int in a side-to-side or circular pattern.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the hardinge fem cement rester would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : hcp is reporting the setting instrument for the medullary canal stopper broke off when it was unscrewed.No force was used.The approximately 6 cm long fragment could not be removed and had to remain in the patient.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of the returned sample revealed that the hardinge cem restr int was found broken from the shaft, the fracture originated from the number 7 marked on the shaft, indicating that approximately 6 cm from the tip broke off the instrument.No postoperative x-rays were provided to confirm the foreign body left in patient.The shaft breakage is consistent with the user applying excessive leverage/torque on the hardinge cem restr int in a side-to-side or circular pattern.A dimensional inspection was unable to be performed due to post manufacturing damage.A functional test was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the hardinge cem restr int would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : manufacturing records evaluation (mre) was not performed as no lot number was able to be retrieved for this device.
|
| |
|