• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS T4; RADIOIMMUNOASSAY, TOTAL THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS T4; RADIOIMMUNOASSAY, TOTAL THYROXINE Back to Search Results
Catalog Number 09007784190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was acceptable.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys t4 assay results for 3 patient samples on a cobas e 801 analytical unit.The customer questioned the high initial results which prompted them to repeat the samples.For sample 1, the initial t4 result was 24.9 ug/dl with flag.The sample was repeated and the repeat result was 13.9 ug/dl.The sample was repeated a third time the next day and the result was 8.56 ug/dl.For sample 2, the initial t4 result was 24.9 ug/dl with flag.The sample was repeated the next day and the repeat result was 8.47 ug/dl.For sample 3, the initial t4 result was 24.9 ug/dl with flag.The sample was repeated and the repeat result was 8.60 ug/dl.The sample was repeated a third time the next day and the result was 8.63 ug/dl.
 
Additional Manufacturer Narrative
The investigation did not identify a product problem.The root cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS T4
Common Device Name
RADIOIMMUNOASSAY, TOTAL THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18801756
Report Number1823260-2024-00582
Device Sequence Number20672708
Product Code CDX
UDI-Device Identifier07613336171677
UDI-Public07613336171677
Combination Product (Y/N)Y
Initial Reporter CountryBR
PMA/510(K) Number
K961490
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Catalogue Number09007784190
Device Lot Number741920
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date02/28/2024
Supplement Report FDA Received Date04/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-