| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It has been reported that a versacross access solution kit was selected for use for a cryoablation procedure indicated for a case of atrial fibrillation (a fib).During the procedure, the physician mentioned that the mechanical guidewire (mgw) wire got stuck in the versacross sheath and could not be removed.The issue occurred when the physician was inserting the versacross sheath towards the right atrium, prior to transseptal.The guidewire and sheath were removed altogether, and the device was replaced (different device, same model).The procedure was then completed successfully.No patient complications reported.Product is expected to return for analysis.No kinks were noted on the guidewire after removal from the body or prior to use.The patient did not have any implanted devices during this procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental will be filed.
|
| |
|