| Model Number |
EQUINOXE CAGE GLENOID LARGE, BETA |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Naturally Worn (2988); Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377)
|
| Date of Event |
01/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient, initial right shoulder implanted on (b)(6)2014, underwent a revision procedure on (b)(6)2024, approximately 9 years 10 months post the initial procedure.The patient presented with complaints of pain.The glenoid component had migrated from its original position.Significant wear and osteolysis found.Bone grafting and conversion to reverse was performed.There were no issues with the surgery.X-rays were provided.Explanted devices are not available for return.Device images were provided.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitants: 2532751-310-01-53 - equinoxe, humeral head short, 53mm (beta).2719489-300-01-15 - equinoxe, humeral stem primary, press fit 15mm.2770139-314-13-14 - equinoxe cage glenoid large, beta.2840008-300-10-15 - equinoxe replicator plate 1.5mm o/s.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported was likely the result of prosthesis wear.Additional reasons for the reported revision may be migration and inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation.D1: corrected.H6: corrected health effect, medical device, component, and investigation clinical codes.
|
| |
|