Catalog Number UNK_PENTARAY
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 02/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation ablation with a pentaray nav high-density mapping catheter and the patient experienced pericardial effusion that required pericardiocentesis.A pericardial effusion was noticed during the procedure.The patient had a baseline effusion; however, once they had gained access to the left side, the patient's blood pressure dropped and it was confirmed with intracardiac echocardiography (ice) that the pericardial effusion had grown.Pericardiocentesis was performed.The procedure was aborted.The patient was reported to be in stable condition.No ablation had been performed.They had lost view of the transseptal wire on ice while puncturing it and that was believed to be the cause of the injury.A pentaray catheter and an smarttouch sf catheter were inside the patient's body.Additional information was received.The adverse event was discovered during use of biosense webster, inc.Products.The sheath used for transseptal puncture was the baylis versacross wire and sheath.The physician¿s opinion on the cause of the adverse event was that it was procedure related as lost sight of the transseptal needle.Patient stayed an extra night for extended monitoring.
 
Additional Manufacturer Narrative
Additional information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturing record evaluation (mre) cannot be conducted because no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV HIGH-DENSITY MAPPING CATHETER
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18804823
Report Number2029046-2024-00678
Device Sequence Number15266079
Product Code MTD
UDI-Device Identifier10846835009576
UDI-Public10846835009576
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberUNK_PENTARAY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2024
Initial Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BAYLIS VERSACROSS WIRE AND SHEATH; UNK_NGEN RF GENERATOR; UNK_SMART TOUCH BIDIRECTIONAL SF
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient SexFemale
-
-