• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065753106
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the infusion tube which does not come off usually, was found came off and leakage occurred during vitrectomy surgery.The surgery was completed after replacing the product with another tube.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The customer returned one pak.The sample was visually inspected and it was confirmed that a weak weld between the top and base of the auto infusion valve which can result in the customer's experience.The root cause of the customer's complaint was determined to be related to equipment not maintaining consistent control of the weld process.Contributing factors identified were training issues during production when performing the integrity squeeze check on the component to verify a good weld and when performing maintenance on the equipment.Action was taken to determine root cause and to take appropriate corrective actions for complaints of a similar nature.Quality assurance identified the likely contributors to be related to equipment end of life and will be replaced with a new system.Procedural enhancements were also implemented for better control of process during quality checks and for better training of personnel during production and equipment maintenance.Complaints are reviewed and monitored at regular intervals for adverse trends.No further action has been identified for this reported event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18805620
MDR Text Key336523665
Report Number1644019-2024-00366
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657531066
UDI-Public00380657531066
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number8065753106
Device Lot Number14MKLV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received03/20/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-