| Catalog Number |
1012632-19 |
| Medical Device Problem Codes |
Off-Label Use (1494); Migration (4003)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the procedure was to treat a stenosed lesion in the brachiocephalic artery.After the lesion was pre-dilated with a non-abbott balloon catheter, the 9.0x19mm omnilink elite stent delivery system (sds) was inflated to 11 atmospheres (atm) for 10 seconds.When the sds was attempted to be removed, the stent migrated proximal to the target lesion.A non-abbott balloon catheter was used to over dilate the migrated stent to complete the procedure.There was no clinically significant delay in the procedure and no adverse patient sequela.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It was reported the procedure was to treat a stenosed lesion in the brachiocephalic artery.It should be noted that the omnilink elite instructions for use (ifu) states: the omnilink elite stent system is indicated for the treatment of atherosclerotic iliac artery lesions.It is unknown the ifu deviation contributed to the reported event.The investigation was unable to determine a conclusive cause for the reported stent migration.The reported treatment and foreign body in patient appear to be related to operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Search Alerts/Recalls
|