| Model Number |
APOLLORF HOOK, NON-ASPIRATING 90° |
| Medical Device Problem Codes |
Arcing of Electrodes (2289); Arcing (2583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy the plasma field formed, not only at the hook probe, but also at the entire tip of the probe.The surgeon then stopped using the probe for the operation because it was too risky for him.There was no harm for patient, operator or third party reported.The surgery was finished successfully.It was not necessary to switch the surgical technique or do a second surgery.Update nroe 13-feb-2024: further information revealed that the surgeon used the same device to finish the surgery, knowing that the plasma field was larger than usual.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is not confirmed.One unpackaged ar-9825 apollo batch: 2206130 was returned for investigation.Visual inspection noted no problems with the device.The device was tested before the memory was cleared and the screen showed, ¿probe already used/replace probe.¿ after the memory was cleared, functional testing was performed with saline water and the probe worked as intended.The plasma field was observed to only form on the hook of the device.No problem found.
|
| |
|
Search Alerts/Recalls
|