It was reported that they had 3 balloons of foley catheter burst inside of the patient.Per additional information received via customer follow up on (b)(6) 2024, they stated that the catheter makes it a couple of weeks before failing.At this time there was no missing balloon pieces, but testing was still on going to find the cause and make sure there was no pieces left inside.Patient experienced extreme pain and mental stress due to the usage of device.No medical intervention was reported.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event is confirmed - cause unknown.Visual: received one (1) used 2-way foley catheter without original packaging.Visual inspection noted the balloon had burst measuring 0.4985".Further inspection noted the balloon had no missing pieces.This is out of specification per inspection procedure which states, "balloon must not be torn".A potential root cause for this failure could be pinhole in balloon or cuff.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: ¿warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Sterile unless package is opened or damaged.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Catheters should be replaced in accordance with the cdc guideline ¿guideline for prevention of catheter-associated urinary tract infection¿.At the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter-related adverse effect, the catheter should be replaced.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that they had 3 balloons of foley catheter burst inside of the patient.Per additional information received via customer follow up on 15feb2024, they stated that the catheter makes it a couple of weeks before failing.At this time there was no missing balloon pieces, but testing was still on going to find the cause and make sure there was no pieces left inside.Patient experienced extreme pain and mental stress due to the usage of device.No medical intervention was reported.
|