| Catalog Number |
AA6310 |
| Medical Device Problem Code |
Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the available information upon checking the foley balloon it was difficult to inflate and the water inserted into the balloon could not be withdrawn.Catheter was not used.
|
| |
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaints and we found another complaint on the lot n° 8559622 (b)(4).The aa63101002 lot 8559622 has been packaged with intermediate products ref (b)(4) lot 7983323 in may 2023 and the expiry date is may 2028.The aa631080 lot 7983323 has been manufactured by our tunisian site.(informed) checking the quality databases revealed no anomaly in connection with the described defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Unfortunately, no sample was received from the customer, and we cannot go further than the documentary investigation which reveals one non-conformity recorded during production.Checking the quality databases revealed a corrective and preventive action possibly in relation with the described issue.The trending for balloon bursting is specifically monitored.A similar case study based on item number aa6310 and defect balloon over the last four years: two similar cases were found.A risk management file evaluation was performed, and concluded that residual risks are adequately controlled and reduced as far as possible, and the residual risks associated with the use of the product are acceptable when weighed against the benefits to the patient/user.It is concluded that the risks identified are still acceptable and considered as safe.
|
| |
|
Search Alerts/Recalls
|