Model Number SVS-V6-00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 01/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Three spiration valves were placed in the patient.Three event reports were filed separately, one for each device.This is report one of three for the related pneumothorax event.As part of the event follow-up to gather additional information regarding the pneumothorax event, gyrus learned that the patient died approximately 10 days post procedure.The cause of death is unknown.No correlation between the death and the adverse event was reported.Pneumothorax is the most common device related serious adverse event that is associated with the spiration valve system.In the emprove study, the incidence of serious pneumothorax was 14.2%, with over 75% of the serious pneumothorax events occurring within the first 3 days post-procedure.Early-onset pneumothorax in the treatment group likely resulted from lung conformation changes due to acute reduction in lung volume by valve therapy, triggering rapid expansion of the ipsilateral non-targeted lobe leading to pneumothorax.(criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration® valve system (emprove): a multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).However, it should be noted that pneumothorax events are also recognized as an indicator of successful target lobe occlusion and when managed according to published expert guidelines (valipour a, slebos dj, de oliveira hg, et al.Expert statement: pneumothorax associated with endoscopic valve therapy for emphysema -- potential mechanisms, treatment algorithm and case examples.Respiration 2004 2014;87(6): 513-521.Doi:10.1159/000360642), subjects with pneumothorax events experienced clinical benefits similar to that in subjects without pneumothorax events (criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration valve system (emprove).A multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).The reported event aligns with the experience in the emprove clinical trial and is an expected adverse event associated with the spiration valve system.H3 other text : valve remained in place.
 
Event or Problem Description
The patient underwent valve placement procedure for the treatment of emphysema on (b)(6) 2024.The physician placed three spiration valves (1x svs-v6-00, 1x svs-v7-00, and 1x svs-v9-00).The patient experienced a pneumothorax immediately post placement procedure after extubating.The patient was re-intubated, and a follow up chest x-ray was taken.A chest tube was then placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE IN CARTRIDGE, 6MM
Common Device Name
SPIRATION VALVE
Manufacturer (Section D)
GYRUS ACMI, INC.
6675 185th ave ne
redmond WA 98052
Manufacturer (Section G)
GYRUS ACMI, INC.
6675 185th ave ne
redmond WA 98052
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18809805
Report Number3004450998-2024-00009
Device Sequence Number17949684
Product Code NJK
UDI-Device Identifier00896506002293
UDI-Public896506002293
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P180007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSVS-V6-00
Device Lot NumberWS236407-01
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2024
Initial Report FDA Received Date02/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-