• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR Back to Search Results
Model Number 186-1001-RFB
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, when the electromedical technician was checking the unit, it was not ringing.However, the unit had visual alarms.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the speaker connector was damaged.Functionally, the device was successfully tested for data transfer to usb drive.The issue was resolved by replacing all faulty parts.It was reported that the sound alarm does not work.The reported issue was confirmed.The most likely cause was traced to a component failure.The evaluation detected an unreported condition: of battery cover mounting stud was broken and hypertronics connector chassis loose.The most likely cause for the additional condition is were traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Common Device Name
INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18809840
Report Number2936999-2024-00399
Device Sequence Number7248798
Product Code OLW
UDI-Device Identifier10884521130173
UDI-Public10884521130173
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K072286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186-1001-RFB
Device Catalogue Number186-1001-RFB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by Manufacturer04/15/2024
Initial Report FDA Received Date02/29/2024
Supplement Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-