EXACTECH, INC. EQUINOXE CAGE GLENOID MEDIUM, BETA; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
314-13-13 |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Medical device site infection (4845); Aseptic loosening (4883)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(h3) pending evaluation (d10) concomitant device(s): 300-30-10 - equinoxe preserve stem 10mm: 5582373 300-50-45 - 4.5mm short rep plate: 5416244 310-01-47 - equinoxe, humeral head short, 47mm (beta): 5193690 315-35-00 - glnd kwire: 5665042 300-20-02 - equinox square torque define screw drive kit: 5442999.
|
| |
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, the 56-year-old male patient had an initial right tsa on (b)(6) 2018 and presented with polyethylene wear (with or without osteolysis), approximately 1 month(s) and 5 year(s) post-operatively on 15-nov-2023.Reviewing x-rays and patient has poly wear.Patient also has increase in pain.The outcome of this event is considered continuing, with the action taken of other - monitoring with x-ray as patient did not want to proceed with revision intervention immediately.The case report form indicates that this event is definitely related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
|
| |
|
Event or Problem Description
|
|
Approximately 5 year(s) and 20 day(s) post-operatively of initial implantation of a right tsa, it was reported via clinical trial study that the patient presented with polyethylene wear (with or without osteolysis).X-rays and monitoring showed and lead to the patient seeking a standard total with preserve stem revision 10 months later, in which the stem and glenoid were replaced.During the surgery, it was reported deep infection with aseptic glenoid loosening.The outcome of this event is now considered resolved, and the case report form indicates that this event is definitely related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
(d10) concomitants: 300-30-10 - equinoxe preserve stem 10mm: 5582373; 300-50-45 - 4.5mm short rep plate: 5416244; 310-01-47 - equinoxe, humeral head short, 47mm (beta): 5193690; 315-35-00 - glnd kwire: 5665042; 300-20-02 - equinox square torque define screw drive kit: 5442999.This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b2, d6b, g1-contact first/last name, h6 - clinical codes, medical device problem code, component code, type of investigation, investigation findings, and investigation conclusions.The following sections were corrected: b5, d10, h6 - impact codes.No device was returned for evaluation; the reported event was confirmed by review of photographs and radiograph images of the device.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6: health effect - clinical code, health effect - impact code, medical device problem code, component code, type of investigation, investigation findings, investigation conclusions multiple mdr reports were filed for this event, please see associated reports: 1038671-2025-01893.The pain and subsequent revision reported may be the result of the aseptic glenoid loosening, prosthesis wear, and infection as reported.However, the failures cannot be confirmed as no relevant clinical information, images, or radiographs were provided.Potential contributions of manufacturing, patient, or user-related issues to the event cannot be determined from the reported information.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|