| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Fall (1848); Bone Fracture(s) (1870); Hemorrhage/Blood Loss/Bleeding (1888); Pain (1994); Loss of Range of Motion (2032); Osteolysis (2377); Muscle/Tendon Damage (4532); Insufficient Information (4580)
|
| Date of Event |
07/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 650-1065 biolox taper sleeve ¿ 835650 13-104060 ¿ m/h shell ¿ 793250 ep-105995 ¿ epoly liner ¿ 346300 11-301412 ¿ arcos distal stem ¿ 620880 unknown arcos cone ¿ unknown part and lot the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2024 - 00078.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent a revision surgery approximately 7 years post implantation.Subsequently, the patient was revised a second time approximately 7 years later due to unknown reasons.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.D10 : ref: (b)(4), lot: 170660 arcos brch sz b hi 60mm.Ref: (b)(4), lot: 158390 arcos 12x115mm cyl dist.Ref: (b)(4), lot: 793250 m/h radial solid/apx shl 60mm.Ref: (b)(4), lot: 346300 epoly 36mm rlc lnr mrom sz25 ze 25.Ref: (b)(4), lot: 835650 cer option type 1 tpr sleve -3.Cerament bone filler, item: unknown, lot: unknown.Unknown cement, item: unknown, lot: unknown.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported by legal allegations that a patient had an initial left total hip arthroplasty approximately sixteen years ago followed by a revision about seven years post implantation due metal related pathology.Subsequently, the patient had a chronic trochanteric avulsion fracture and developed pain, stiffness, range of motion issues, and a fall that led to ongoing diagnostic tests with suspicion of alval.Approximately two years ago, the patient underwent a stage i revision for infection.During the procedure, a small fracture of the anterior cortex occurred from the wedging of the osteotome, the patient experienced blood loss, all implants were removed, and a competitor cement spacer was implanted.Stage ii revision followed one and a half months later with competitor product implanted.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records, raw material certs and sterile certs confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Medical records were provided and reviewed by a health care professional.The review identified that the patient experienced chronic recurrent lateral hip pain and revision was conducted for infection.Further, the root cause of the event of infection is not device related.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, implanted products are not identified as the source or contributing to the reported infection.A definitive root cause could not be determined for the pain.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|