| Catalog Number |
AK-05502 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The report states, "i've encountered confirmed loss of resistance, but the epidural catheter would not pass into the patient".Additional information received states that "patient last week had an inadvertent dural puncture.The epidural catheter would not advance, and so the touhy needle was advanced".The patient status is reported as "back to normal".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
The report states, "i've encountered confirmed loss of resistance, but the epidural catheter would not pass into the patient".Additional information received states that "patient last week had an inadvertent dural puncture.The epidural catheter would not advance, and so the touhy needle was advanced".The patient status is reported as "back to normal".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter and epidural needle with no relevant findings.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|