| Model Number |
TFL-SLS |
| Medical Device Problem Codes |
Break (1069); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, that the laser system had broke at the connector flexible part and sparked.The issue was found during the therapeutic urethroscopy with laser lithotripsy procedure which was completed with a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d9, should have been marked yes only.The device was not returned to olympus for physical evaluation.The device was evaluated onsite by an olympus field service engineer.The customer's reportable malfunction was not reproduced.The device was within performance specifications.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause for the event could not be determined as the event was not confirmed.Olympus will continue to monitor field performance for this device.
|
| |
|