| Catalog Number |
33680001 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Insufficient Device Problem Information (3190); Migration (4003)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Inadequate Osseointegration (2646); Insufficient Information (4580)
|
| Date of Event |
02/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.H3: other text: device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, based on available medical records and assessment of health care professionals.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed: after reviewing hcp opinion and surgeon¿s feedback we can conclude that the tibial tray has been cement augmented for significant anterolateral bone loss.The tibia shows clear radiolucence and some subsidence.Tibial component could be loosening and a migrated.Pe seems attached with components, no sign of breakage or separation.The talar component does not show significant radiolucence or cysts.Hence, no loosening or migration can be confirmed for talar component.Based on investigation, the root cause was attributed to a patient related issue.The failure is detected by the radiolucence and cysts around tibial component.A screw was also observed while reviewing ct scans without any information about its implantation whether screw inserted during implantation or before implantation.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|