• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804300-15
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6: 2017 device code clarifier- excessive force.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified and mildly tortuous left anterior descending coronary artery (lad).A 3.0x15mm xience skypoint stent delivery system (sds) was advanced into the guide liner (guide extension 5f) however, the delivery system became stuck.When attempting to pull out the sds, resistance was met and force was applied.The shaft broke and at a point outside of the y-connector; therefore, everything was removed together.Another 2.5x15mm skypoint was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported shaft separation was confirmed.The reported difficulty to advance and difficulty to remove could not be replicated in a testing environment as it was related to operational context of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that when attempting to pull out the sds, resistance was met and force was applied.The shaft broke and at a point outside of the y-connector; therefore, everything was removed together.It should be noted that the xience skypoint everolimus eluting coronary stent system instruction for use states: if unusual resistance is felt before the stent exits the guiding catheter, do not force passage.Maintain guide wire placement across the lesion and remove the delivery system and guiding catheter as a single unit.In this case, the instructions for use deviation related to the use of force does not appear to have caused the reported difficulties.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18813321
Report Number2024168-2024-02627
Device Sequence Number4320274
Product Code NIQ
UDI-Device Identifier08717648233234
UDI-Public08717648233234
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1804300-15
Device Lot Number3101641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date02/29/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE LINER. 5F GUIDE EXTENSION.
Patient Age85 YR
Patient SexFemale
Patient Weight49 KG
-
-