MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK EXCELSIOR SL-10 PRE-SHAPED 45 150CM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK EXCELSIOR SL-10 PRE-SHAPED 45 150CM; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number M0031681900
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is report 2 of 4.Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection the returned microcatheter was found to be cut.The operator cut it on the table after the procedure and only part of the catheter was returned.A deployed stent was found inside the catheter shaft.The hub was found to be intact.During the catheter shaft friction functional test the device failed.The device could not be flushed due to friction.A patency mandrel was advanced from the distal end; however, it stuck 10 cm from proximal end.The catheter shaft was cut where the mandrel stuck, and the deployed stent was found inside.On removal of the stent the patency mandrel was advanced through the cut microcatheter shaft and polytetrafluoroethylene (ptfe) and green material was removed.The reported event is covered in the device direction for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported defect was confirmed based on analysis of the device.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information provided by the customer indicated that the device was prepared for use as per the directions for use, there was no damage noted to the packaging prior to opening the packaging, the device confirmed to be in good condition during preparation/prior to use on the patient and continuous flush was set up and maintained throughout the clinical procedure.The catheter shaft was returned cut and only part of the catheter was retuned.The patency mandrel was advanced through the catheter and the concurrent stent was removed from the distal part of the cut shaft.The stent was found to be deformed.The catheter device could not be flushed after the stent was removed.The catheter was cut again, green material and ptfe inner lining were removed from the shaft.The green material was most likely from the guidewire used in the procedure.There was no allegation against the guidewire used in the procedure.The most likely cause for the ptfe peeling from the guidewire is the guidewire may have been backloaded through the guidewire introducer to load into the microcatheter hub.The catheter ptfe inner lining was most likely was damaged when friction was felt advancing the stent.The reported and analyzed event of catheter shaft friction as well as the as analyzed damage of the catheter ptfe inner lining peeling and the device difficult to flush will be assigned procedural factors as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and anatomical factors during use, the product performance was limited.
 
Event or Problem Description
The device was returned for analysis and the investigation of the device revealed that the catheter (subject device) ptfe coating was peeling.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCELSIOR SL-10 PRE-SHAPED 45 150CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key18813340
Report Number3008881809-2024-00079
Device Sequence Number4324583
Product Code DQY
UDI-Device Identifier04546540688224
UDI-Public04546540688224
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042568
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Catalogue NumberM0031681900
Device Lot Number24146874
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2024
Initial Report FDA Received Date02/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient RaceAsian
-
-