Catalog Number 71665014
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during surgery the portion of the set screw driver which passes through the joint fell apart.All pieces were visualized and retrieved from the patient.Surgeon confirmed that no pieces were left via fluroscopy.It is unknown if there was any delay.No patient injury reported.
 
Additional Manufacturer Narrative
H6: the device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
G4: date received by manufacturer.New information was received, with a confirmation of the aware date.Sales representative was not present during surgery, and no smith and nephew employee was aware of this event until the medwatch mas received.Therefore, our aware date is february 28th, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET SCREW DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18813632
Report Number1020279-2024-00460
Device Sequence Number19914738
Product Code HXX
UDI-Device Identifier03596010518989
UDI-Public03596010518989
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71665014
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by Manufacturer02/28/2024
04/17/2024
Initial Report FDA Received Date02/29/2024
Supplement Report FDA Received Date03/05/2024
04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-