Model Number FG- 36045-001
Medical Device Problem Codes Device-Device Incompatibility (2919); Compatibility Problem (2960)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The root cause of the incident is unknown at the time of the initial report.The complaint device will not be returned for an evaluation.The dhr review, which examines the dhr, ncrs, and ecos will be done as part of the root cause investigation.
 
Event or Problem Description
The customer reported as follows: catapult sheath had a hard time entering the 24f impella sheath.Device/procedure outcome: procedure - no information.As reported device codes: 1307: difficult to insert.As reported patient codes: 9398: no clinical signs, symptoms or conditions.Patient outcome: f24: insufficient information.
 
Additional Manufacturer Narrative
The complaint device was not returned for an evaluation.The complaint description is also not clear because the used impella sheath is intended to be used with other devices: more than one device at the same time as specified in impella® procedure with single-access technique procedural steps according to abiomeds brochure "impella procedure with single-acess teechnique.It could be one reason that the appropriate compatibility was not considered.However, no detailed information was received for the medical devices used in the operation other than 24f impella sheath.The hard insertion would be due to improper surface property.However, catapult sheath has hydrophilic coating that helps reducing frictions during the inserion if it is wetted.According to requested information by customer, the preparations before insertion of sheath were completed properly.According to the investigations done on batch record, ncrs, and ecos related to the complained lot, there is no event or deviation that can cause this defect failure mode.Root cause is undeterminable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATAPULT GUIDE SHEATH, 6F, 45CM, STRAIGHT, HYDROPHILIC COATED
Common Device Name
INTRODUCER SHEATH
Manufacturer (Section D)
CONTRACT MEDICAL INTERNATIONAL GMBH
lauensteiner strasse 37
dresden, 01277
GM  01277
Manufacturer (Section G)
CONTRACT MEDICAL INTERNATIONAL GMBH
lauensteiner strasse 37
dresden 01277
Manufacturer Contact
juliana nuernberger
lauensteiner strasse 37
dresden, 01277
GM   01277
MDR Report Key18814880
Report Number3003637635-2024-00006
Device Sequence Number8307813
Product Code DYB
UDI-Device Identifier04251244502096
UDI-Public04251244502096
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K221914
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2026
Device Model NumberFG- 36045-001
Device Catalogue NumberUS-36045-F-ST-H
Device Lot Number752276
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-