EXACTECH, INC. LOGIC CR TIB INSERT SLOPE++, SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-49-4011 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Synovitis (2094); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitants: 4619893 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t; 5033622 02-010-03-0340 - logic cr femoral cem, right, sz 4; 5083337 200-02-32 - three peg patella 32mm.These devices are used for treatment not for diagnosis.Pending investigation.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a right total knee arthroplasty on (b)(6) 2018, and then experienced revision surgical procedure on (b)(6) 2023 approximately 5 years and 1 month after initial implant due to failure of right knee with recalled polyethylene and severe osteolysis per revision operative report received.No images were provided.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b1, b2 the following sections were corrected: g1, g2, h6 mdr section codes updated/corrected: b, c, d, e the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|