MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; CHAIR, ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; CHAIR, ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number GL358XL/LC358XL
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Bruise/Contusion (1754)
Date of Event 02/15/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
 
Event or Problem Description
Provider alleges the consumer alleges he was reclined in the chair and the remote's lights allegedly flashed on and off he did not think any thing of it but then while talking with his brother; sparks allegedly started to come out of the remote as he was repositioning the chair to be more comfortable, consumer alleges he was in a panic and his brother told him to roll or slid out of it which he slid down and rolled off it.
 
Event or Problem Description
Provider alleges the consumer alleges he was reclined in the chair and the remotes lights allegedly flashed on and off he did not think any thing of it but then while talking with his brother; sparks allegedly started to come out of the remote as he was repositioning the chair to be more comfortable, consumer alleges he was in a panic and his brother told him to roll or slid out of it which he slid down and rolled off it.
 
Additional Manufacturer Narrative
Damaged hand control harness.The product literature contains instructions to periodically check the hand control for visible damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
CHAIR, ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18643
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york avenue
duryea, PA 18344
8008008586
MDR Report Key18815257
Report Number2530130-2024-00334
Device Sequence Number20900808
Product Code INO
UDI-Device Identifier00606509400299
UDI-Public00606509400299
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K070950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberGL358XL/LC358XL
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2024
Supplement Date Received by Manufacturer02/16/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-