• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 0170SI12
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that they found black particle inside the foley catheters.Impacted product as follows - they were found prior to patient use and there was no patient impact.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was available for evaluation.Root cause could not be identified.Although a root cause could not be definitively identified, a potential root cause for this type of failure could be ¿not following the production areas cleaning procedure".However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Sterile unless package is opened or damaged.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Catheters should be replaced in accordance with the cdc guideline ¿guideline for prevention of catheter-associated urinary tract infection¿.At the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter-related adverse effect, the catheter should be replaced.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities; 5cc balloon: use 10cc sterile water; 30cc balloon: use 35cc sterile water; do not exceed recommended capacities." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Event or Problem Description
It was reported that they found black particle inside the foley catheters.Impacted product as follows - (material #0170si12 and batch# nggq1259), (material #165818 and batch# nghv4431), (material #165816 and batch# nghs0849).They were found prior to patient use and there was no patient impact.Per customer follow up received on 29mar2024, it was reported that the affected foleys were returned under ref #(b)(4) a few weeks ago.The lots listed in the mentioned complaint were the only ones affect.As stated they were found before patient contact or use.
 
Event or Problem Description
It was reported that they found black particle inside the foley catheters.Impacted product as follows - (material #0170si12 and batch# nggq1259), (material #165818 and batch# nghv4431), (material #165816 and batch# nghs0849).They were found prior to patient use and there was no patient impact.Per customer follow up received on 29mar2024, it was reported that the affected foleys were returned under ref # (b)(4).A few weeks ago.The lots listed in the mentioned complaint were the only ones affect.As stated, they were found before patient contact or use.
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted received 1 all silicone foley catheter.The original package was opened but the blue sleeved was sealed.Per visual inspection it was noted that there was a brown/black loose particle inside the blue sleeve.The blue sleeve was opened and a foreign material was found measuring 5 sq mm.Based on the physical sample received the reported event is confirmed and does not meet specifications per ip7600285 rev 33 which states "loose foreign matter, engraved and spots >1/32¿ is not allowed".No actions can be taken at this time since a root cause was not identified.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Sterile unless package is opened or damaged.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Catheters should be replaced in accordance with the cdc guideline ¿guideline for prevention of catheter-associated urinary tract infection¿.At the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter-related adverse effect, the catheter should be replaced.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.Correction: d, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18815864
Report Number1018233-2024-01082
Device Sequence Number17215741
Product Code MJC
UDI-Device Identifier00801741025327
UDI-Public(01)00801741025327(11)220408(17)260131(10)NGGQ1259
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K984136
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Catalogue Number0170SI12
Device Lot NumberNGGQ1259
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2024
Supplement Date Received by Manufacturer04/30/2024
02/04/2025
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date05/02/2024
02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-