| Model Number |
CORING REAMER, 10MM WITH COLLARED PIN |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by an arthrex subsidiary employee via sems (b)(4) that the surgeon has used the coring reamer, ar-1224s, in many cases and is familiar with the technique.He followed the procedure to drill the tibial tunnel and upon exiting into joint, the tip of the coring reamer expanded and flowered open.There are multiple splits in the reamer near the tip.He was able to remove the drill and no remains were inside the joint.The issue was noticed after the device exited the cavity.There was no adverse effect on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional information from the field, arthrex concluded the cause of the reported failure.The most likely cause of the reported failure is user error.Excess drilling force, in conjunction with any angulation changes of the drill during reaming, may result in coring reamer deviation or failure of the device.The complaint allegation was not confirmed.The device was not received for evaluation, and no evidence of the failure was provided.
|
| |
|
Search Alerts/Recalls
|