Catalog Number 368659
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D.2.Medical device type: one additional code applies, jka g5: pma/510(k)#: one additional code applies; k220212 h.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that during the use of bd vacutainer® push button blood collection set with pre-attached holder, the blood enters the tubing really fast when the needle was inserted into the patient.Then when the vacutainer tube was connected, it shoots the blood in really quickly and loses all pressure; it stops working.No patient impact reported.
 
Event or Problem Description
It was reported that during the use of bd vacutainer® push button blood collection set with pre-attached holder, the blood enters the tubing really fast when the needle was inserted into the patient.Then when the vacutainer tube was connected, it shoots the blood in really quickly and loses all pressure; it stops working.No patient impact reported.
 
Additional Manufacturer Narrative
The following information has been updated with corrected and/or additional information: d.9.Device available for eval? yes.D.9.Returned to manufacturer on: 06-mar-2024.H.6.Investigation summary: "material #: 368659.Lot/batch #: 2297392.Bd received 1 sample and 1 photo for investigation.The photo was reviewed, and it shows the unit packaging of the device, confirming the lot number.Additionally, the customer sample along with 30 retention samples from bd inventory were evaluated by functional draw testing and the indicated failure mode for insufficient flow with the incident lot was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode insufficient blood flow.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18817530
Report Number1024879-2024-00189
Device Sequence Number4324427
Product Code FPA
UDI-Device Identifier50382903686594
UDI-Public(01)50382903686594
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K030573
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2024
Device Catalogue Number368659
Device Lot Number2297392
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2024
Supplement Date Received by Manufacturer03/08/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-