| Model Number |
1119.0011 |
| Medical Device Problem Code |
Expulsion (2933)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being resubmitted for an incident that occurred earlier.The devices were not available for return.Imaging provided shows one locking cap visually out of the screw head at the top of the construct.Possible causes that may have contributed to the loosening include intra-operative tightening conditions, excessive post-operative loading, or unknown implant related causes.The exact cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was done to replace 4 locking caps which had come loose post-operatively.
|
| |
|