|
D10 concomitant products: 444sp02014, 444sp02014 14ch adt ty-care exel x10 (lot#22j0616fax); 444sp02014, 444sp02014 14ch adt ty-care exel x10 (lot#unknown) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter, an air leak was heard in the intubation tube.Observation of the device revealed that the closed system was broken on the respirator side.The system was changed immediately and five minutes after fitting the new closed system, the same problem was noted with a leak.When the closed system was replaced with a hypoxemic connector, the closed system broke, crumbling into the ventilator circuit and making it impossible to reconnect the patient and ventilate.The doctor had to intervene urgently, and the patient was ventilated using a bavu (self-filling balloon with one-way valve).There was no reported patient outcome.
|