• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS LIMB SALVAGE SYSTEM; MALE-FEMALE MIDSECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS LIMB SALVAGE SYSTEM; MALE-FEMALE MIDSECTION Back to Search Results
Lot Number 1935698
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 02/18/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
All inspection and manufacturing data was reviewed for the device lots listed in the table above; the lots were found to be conforming to specifications and no manufacturing abnormalities were observed.As detailed in section 3.5, the ifu and surgical technique documentation identifies elements such as patient contraindications, patient selection factors, surgical procedures/techniques and other precautions/conditions that may contribute to adverse effects.The root cause of this complaint was not determined.Based on the review of device history records, the investigation concluded that the root cause of the alleged infection was not related to the design, manufacture, and/or sterilization of previously implanted eleos implants.The following mdrs are related to this adverse event: 3013450937-2024-00035 3013450937-2024-00036 3013450937-2024-00037 3013450937-2024-00038 3013450937-2024-00039.
 
Event or Problem Description
It was reported by t.Stamper, an onkos sales representative, that a 62-year-old female patient with an eleos distal femur replacement underwent a revision surgery due to an alleged infection on (b)(6) 2024.The surgeon suspected an infection because the surgical wound never fully healed.The primary surgery took place on (b)(6) 2023.The patient was reportedly an oncology patient.All eleos explanted implants were replaced with new implants during the washout revision surgery.These devices were not returned for evaluation.All other eleos devices implanted during the primary surgery remain implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS LIMB SALVAGE SYSTEM
Common Device Name
MALE-FEMALE MIDSECTION
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer Contact
yara neurauter
77 east halsey road
parsippany, NJ 07054
MDR Report Key18818119
Report Number3013450937-2024-00038
Device Sequence Number19961378
Product Code KRO
UDI-Device IdentifierB27825001050E0
UDI-PublicB27825001050E0
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot Number1935698
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Initial Report FDA Received Date03/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISTAL FEMUR; DISTAL FEMUR AXIAL PIN; MALE-FEMALE MIDSECTION; TIBIAL HINGE; TIBIAL POLY SPACER
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-