| Model Number |
STEP DRL,2.4MM/3.0MM TISSTAK/BIO-S-TAK |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 02/15/2024, it was reported by an arthrex subsidiary employee via email that an ar-1250lt step drill for 3.0mm bio-suturetak broke and got stuck in the patient's bone.The broken fragments were not retrieved and remain in the patient.This was discovered during a procedure on (b)(6) 2024.No further information was provided.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misaligned insertion; prying/leveraging the device during insertion.
|
| |
|