Model Number K10001141
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
It was reported that the pump head does not close over the tubing.It was also reported that the pump head had never been replaced and was in bad shape.The procedure to be performed was diagnostic in nature and was completed using the same equipment.There were no reports of patient harm.
 
Additional Manufacturer Narrative
Updated fields: g3, h2, h4, h6, h10 this report is being supplemented to provide additional information based on the approved final investigation.The device was not returned to olympus for evaluation, as customer advised the device will be retained.The customer was advised the pump head be replaced annually.The most probable cause of the complaint is traced to component failure, which is expected or random component failure without any design or manufacturing issue.The customer was provided replacement part information and educated on device replacement guidelines.A device history review of the current product was conducted, and no problems were found.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
It was reported that the pump head does not close over the tubing.It was also reported that the pump head had never been replaced and was in bad shape.The procedure to be performed was diagnostic in nature and was completed using the same equipment.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUSHING PUMP OFP-2 (US)
Common Device Name
OLYMPUS FLUSHING PUMP
Manufacturer (Section D)
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
keymed house, stock road
southend on sea, essex SS25Q H
UK  SS25QH
Manufacturer (Section G)
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
keymed house, stock road
southend on sea, essex SS25Q H
UK   SS25QH
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18818251
Report Number9611174-2024-01020
Device Sequence Number15549353
Product Code FEQ
UDI-Device Identifier15019778003191
UDI-Public15019778003191
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K100899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberK10001141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer03/01/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date03/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/19/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-