• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00558490
Medical Device Problem Codes Human-Device Interface Problem (2949); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a040609 captures the reportable event of balloon tip bent.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the colon during a colonoscopy procedure performed on (b)(6) 2024.During the procedure, a photo of the device was provided by the customer and showed that the tip of the balloon catheter was bent.The procedure was completed with the original device.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the colon during a colonoscopy procedure performed on (b)(6) 2024.During the procedure, a photo of the device was provided by the customer and showed that the tip of the balloon catheter was bent.The procedure was completed with the original device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a040609 captures the reportable event of balloon tip bent.Block h10: investigation results the returned cre wireguided dilatation balloon was analyzed, and a visual and media inspection found that the tip of the guidewire was bent.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon tip bent was confirmed.The results of the analysis performed on the returned device found that the tip of the guidewire was bent.The tip bent found is likely to have occurred due to an excessive manipulation of the device without enough care during unpacking of the device.Therefore, the most probable root cause is an adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18818665
Report Number3005099803-2024-00799
Device Sequence Number8383294
Product Code FDT
UDI-Device Identifier08714729339427
UDI-Public08714729339427
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/23/2025
Device Model NumberM00558490
Device Catalogue Number5849
Device Lot Number0030404909
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by Manufacturer03/08/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date04/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-