| Catalog Number |
03110664180 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lithium electrode lot number and expiration date were requested, but not provided.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for two patient samples tested with the lithium electrode (li+) on a 9180 electrolyte analyzer.The first sample initially resulted in a lithium value of 0.23 mmol/l and it repeated as 0.29 mmol/l.The sample was repeated on (b)(6) 2024, resulting in a value of 0.69 mmol/l.The second sample initially resulted in a lithium value of 0.53 mmol/l on (b)(6) 2024.The sample was repeated twice on (b)(6) 2024, resulting in lithium values of 1.79 mmol/l and 1.53 mmol/l.
|
| |
|
Additional Manufacturer Narrative
|
|
It was determined that the lithium electrode was installed on the instrument on 21-oct-2023.The on-board use time for the lithium electrode is 13 weeks.The in use time for the electrode was exceeded.It was also noted that the customer did not perform regular deproteinizing of the instrument.Per product labeling: "incorrect results may lead to errors in diagnosis, posing danger to the patient.Do not use reagents or consumables that have exceeded their expiry date, otherwise you may obtain inaccurate data." the investigation did not identify a product issue.
|
| |
|
Additional Manufacturer Narrative
|
|
G1 contact office-manufacturing site was updated.
|
| |
|