INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1114
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 02/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information received indicating the following: 1.When placing the skull clamp, if you were to draw an imaginary line between the single pin and the rocker arm swivel point, did that imaginary line pass through an axial center of the skull? yes.2.Did each pin engage the cranium in a perpendicular approach? yes.3.Was there a delay in surgery due to product problem? yes, 30 min.4.Was there any patient adverse consequence because of the delay? no.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The mayfield infinity skull clamp (a1114) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The definite root cause cannot be identified as the device was not returned.However, a probable root cause for the reported complaint is improper or suboptimal placement of the skull clamp on the patient.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
A facility reported that during patient initial positioning for a "posterior cervical" case slippage occurred while patient was prone.The patient had to be flipped into supine position and scalp laceration was repaired with sutures.Patient was able to be re-pinned and surgery was completed.No information on surgical delay was provided.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Common Device Name
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18819520
Report Number3004608878-2024-00028
Device Sequence Number18210617
Product Code HBL
UDI-Device Identifier10381780253549
UDI-Public10381780253549
Combination Product (Y/N)N
PMA/510(K) Number
K051440
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberA1114
Device Lot Number221
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer03/07/2024
Not provided
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date03/22/2024
04/18/2024
Date Device Manufactured03/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-