| Catalog Number |
A1114 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported that the patient's head slipped out of the mayfield infinity skull clamp (a1114).There was no patient injury and no known surgical delay.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The mayfield infinity skull clamp (a1114) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit showed movement in the lock with rotational and lateral movement and a residue buildup.As a result, new components were added to replace worn internal parts and general cleaning and maintenance were performed.Root cause - the complaint is confirmed via inspection of the unit.The root cause is confirmed routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|