| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Misfocusing (1401)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens -11.00/+4.0/109 (sphere/cylinder/axis), into the patient`s right eye (od) on (b)(6) 2024.The lens was explanted on (b)(6) 2024 due to patient experienced a refractive surprise.The lens was replaced with another same model/length but different diopter lens and the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid, in a specimen cup.Visual inspection found no visible damage to the lens.Dimensional and refractive verification were performed and the lens was found to be in specification.Claim #(b)(4).
|
| |
|