| Catalog Number |
D0146 |
| Medical Device Problem Codes |
Failure to Read Input Signal (1581); Low Readings (2460)
|
| Health Effect - Clinical Code |
Liver Damage/Dysfunction (1954)
|
| Date of Event |
02/03/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Device user reported that the device failed to go into liver on board mode after a donor liver was placed on the device.As a result, the marginal liver was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Service engineer attended site and a service and full functional test was performed.All testing passed and no erratic behavior was seen in the pressure readings and no fault was found with the retained unit.The device data was collected and reviewed.Only the session data from 03 feb 2024 shows an unusual pressure reading and later an erratic random reading during the same perfusion.Device testing and data review suggests that the fault observed was with the disposable set used and not the retained unit.The customer had discarded the disposable set and therefore it is not possible to investigate for a root cause any further.
|
| |
|
Additional Manufacturer Narrative
|
|
During troubleshooting the surgeon mentioned the device was reading -99mmhg.They adjusted tubing and refitted the sensor cable but the issue persisted.The arterial sensor appeared to have an unstable connection.Liver on board mode was achieved eventually, but the arterial pressure readings were irregular.Evaluation of the device is pending.
|
| |
|
Event or Problem Description
|
|
Device user reported that the device failed to go into liver on board mode after a donor liver was placed on the device.As a result, the marginal liver was discarded.
|
| |
|